Leave Your Message

RT-LAMP Probe Master Mix (Lyophilized beads) HYB416

This product contains reaction buffer, Bst DNA polymerase and heat-resistant reverse transcriptase RT-Enzymes Mix, and freeze-dried protectant components. The reaction buffer contains Mg2+, dNTP and other necessary components for amplification.

Cat.No.: HYB416

Specification: 96T/960T/9600T

    This product contains reaction buffer, Bst DNA polymerase and heat-resistant reverse transcriptase RT-Enzymes Mix, and freeze-dried protectant components. The reaction buffer contains Mg2+, dNTP and other necessary components for amplification. The product is in the form of freeze-dried pellets, and only primer probes and templates need to be added when using it.

    Components

    Components

    96 T

    960 T

    9600 T

    RT-LAMP Probe Master Mix (Lyophilized beads)

    8 T/strip×12

    8 T/strip×120

    8 T/strip×1200

     Storage

    Store at 2-8℃, valid for 12 months.

    Instructions

    1. Take out the corresponding number of Lyophilized beads according to the number of tests.

    2. Prepare the reaction system

    Component

    Volume

    RT-LAMP Probe Master Mix (Lyophilized beads)

    One piece

    10×Primer probe Mix*

    5 μL

    Template RNA**

    45 μL

    *10× Primer Mix concentration: 16 μM FIP/BIP, 2 μM F3/B3, 4 μM Loop F/B,4 μM probe;

    **It is recommended to dissolve the nucleic acid template in DEPC water.

    3. Amplification reaction

    Set up the following program on a fluorescent PCR instrument (such as ABI7500, etc.).

    Temperature

    Time

    Cycles

    65℃

    60s

    *Collect FAM fluorescence

    30

     Notes

    1. The reaction temperature can be optimized between 62℃-68℃ according to the primer situation;

    2. The experiment should be operated in a standardized manner, including the preparation of the reaction system, freeze-drying, sample processing and sample addition;

    3. To avoid contamination, it is recommended to prepare the reaction system in a clean bench and add the template in a fume hood in other rooms to avoid false positive interference.

    Leave Your Message